eRHC Calculator PRO
Hemodynamic Calculation Tool
0 ratings
Free
Rating summary
About
eRHC Calculator PRO is a hemodynamic calculator app for qualified healthcare professionals, used during right heart catheterization (RHC) procedures. Here's what it contains:
App Purpose
A clinical calculation and documentation tool for invasive hemodynamic measurements — specifically tied to the ALLAY-HF and ALLAY-HFrEF clinical studies.
Key Features
Patient data management — save/load/share patient records as JSON files, with patient ID, institution, weight/height (kg/lb, cm/in), DOB, and BSA calculation
Multi-stage calculator — timed hemodynamic measurements across stages (rest, exercise, etc.) with vitals, pressures, V-waves, CO via thermodilution or Fick method
Threshold system — configurable alert thresholds with rules (lower/upper bound, range, special PCWP rules)
Time synchronization — reference clock offset for procedure timing
Print/share results — formatted report output
Formula reference — all formulas listed (BSA, mPAP, PCWP, TPG, VO₂, CO, CI, PVR)
Formulas Implemented
BSA: Mosteller (general) and Du Bois (Fick-specific)
mPAP: (sPAP + 2×dPAP) / 3
PCWP from V-Wave: ((V-High − V-Low)/3) + V-Low
TPG: mPAP − PCWP
VO₂: Gender-specific regression equations (Dehmer et al.)
O₂ content: (Sat/100 × Hb × 1.36) + dissolved O₂
CO (Fick): VO₂ / (arteriovenous O₂ difference × 10)
CI: CO / BSA
PVR: (mPAP − PCWP) / CO (in Wood units)
Legal/Info
Developer: Andreas Kuhl, Germany
Fully offline — no data collection, no internet, all data stays on device
Medical disclaimer: intended as a calculation aid only, not for standalone clinical decisions
© 2025-2026 Andreas Kuhl
Clinical References
Citations link to PubMed/journals for Mosteller, Chemla, Rajagopalan, Simonneau, ESC/ERS 2022 PH Guidelines, Du Bois, Dehmer, Severinghaus, Fick, and others.
Show more
What's New in eRHC Calculator PRO
1.6.2
November 2, 2026
• Calculations now follow the CF-0041 worksheet: every displayed intermediate value (mPAP, PCWP, TPG, CO, CI, PVR) is rounded before it is used further, so results match a manual calculation. • Threshold evaluation (colors, OK/not OK markers) now uses the displayed values. • Date of birth can be set to N/A. • Fick CO: choose between study protocol (CD-0077) and clinical practice (LaFarge VO₂), with warnings.
More